Product Specifications

Product series 16 inch 8 cells 16 inch 10 cells 16 inch other
Effective filtration area (EFA) 1.8m2 2.3m2 Optional
Overall dimension (mm) 184 mm 275 mm -
Filter Media Filter Media Cellulose, Filter Aid, Resin or Cellulose
Frame and Supporter Glass Fiber Reinforced Polypropylene
Stainless Steel Band 316L
O-ring Silicone
Particle Challenge Particle Intercept>90%
Max Operation Pressure 3.0 bar @ RT
Max Pressure Differential Forward:2.0bar@RT
Reverse:not recommended
Endurance to Sterilization SIP: 123℃ 30min, 3cycles
Indirect Food Additive The raw materials of all Alidep® Depth Stacked Filter components comply with the requirements forfood additives in FDA 21 CFR 177-182.
Toxicity The raw materials of all Alidep® Depth Stacked Filter components meet the requirements for in vivo biological safety of Class VI plastics in USP <88>.
Bacterial endotoxin Endotoxin level of rinsing liquid is < 0.25 EU/mL, which complies with the requirements for water for injection in Chinese Pharmacopoeia and United States Pharmacopoeia.

 

Conductivity Test Results                                       pH Test Results

 

TOC Test Results                                                   Test Results of Flow Differential Pressure

 

Validation Methods and Results

Conductivity Several filters were randomly selected from different production batches, then rinsed with 100L/m2 pure water for 10-20minutes. The conductivity ranged from 5-8μS/cm.
TOC Several filters were randomly selected from different production batches, then rinsed with 100L/m2 pure water for 10-20minutes. The TOC was about 1mg/L.
pH Several filters were randomly selected from different production batches, then rinsed with 100L/m2 pure water for 10-20minutes. The pH ranged from 6.6-6.8.
Endotoxin Several filters were randomly selected from different production batches, then rinsed with 100L/m2 pure water for 10-20minutes, followed by soaking in pure water for 24 hours. The endotoxin level was < 0.25 EU/mL as indicated by Tachypleus Amebocyte Lysate test.
Endurance to Sterilization Several filters were randomly selected from different production batches, after 3 cycles of SIP at 123℃ for 30min, filters passed air tightness, diffusion flow, hydraulic stress. Flow rate and pressure dropafter SIP met the product quality standard.
Hydraulic Stress Several filters were randomly selected from different production batches, respectively conduct hydraulic stress test at ambient temperature and after 3 cycles of SIP at 123℃ for 30min. High-viscosity solutions was filtered to increase the pressure drop across the filters.Apply forward pulse pressure to the filter at ambient temperature, rinseand conduct diffusion flow. Depth stacked filter can withstand 2.0bar forward pressure at ambient temperature.